Detail MMM* VACUCELL, Oven With Vacuum Drying 111 Litre, MC000926
Vacucell 111 - Standard
Quick, cost-saving, ecological - laboratory oven meeting higher demands
111-liter chamber volume
Working temperature of from + 5° C over ambient temperature up to 200° C
Noiseless operation and soft heating of samples in the chamber; high-quality and safe heating and drying of samples to constant mass
High operation comfort, precise temperature regulation and short times of temperature recovery in the chamber after the door opening
Suitable for perfect drying of thermolabile and oxidation-sensitive materials, complex-shape components, samples drying to constant mass etc. under vacuum
Standard model with a Fuzzy-Logic microprocessor control system; control elements situated on a foil keyboard, process information indicated on a LED display
Wide range of optional equipment and accessories ( e.g. a “ Vacustation” box enabling a selected vacuum pump installation, … )
Output for a printer or PC connection
Validation possibility ( IQ , OQ)
Technical data
Interior of stainless steel material No. 1.4571
volume 111 l
depth 410 mm
width 540 mm
height 480 mm
Tray
number 2 pcs
racks max. 9 pcs
Shelves distance
height 40 mm
Usefull dim. of shelf
width × depths 480× 386 mm
External dimensions ( including door and handle)
depth 640 mm
width 760 mm
heights( incl. foots) 880 mm
Package dimensions ( three layers carton)
depth 730 mm
width 830 mm
height ( incl. palette) 1070 mm
Weight
gross 144, 5 kg
net 130 kg
Working temperature ( regulation start)
from 5 ° C over ambient temp to 200 ° C
Temp. deviations acc. to DIN 12 880 Part 2 from working temp – Al shelves pressure 5– 10 mbar* *
in time to: 0, 4 ± ° C
in space, temperature 100 ° C do: 3 ± ° C
in space, temperature 200 ° C do: < 7 ± ° C
Temp. deviations acc. to DIN 12 880 Part 2 from working temp – ss shelves pressure 5– 10 mbar* *
i time to: 1 ± ° C
in space, temperature 100 ° C do: 11 ± ° C
in space, temperature 200 ° C do: nezm Yeno ± ° C
Time of rise onto 98 % voltage 230 V – ( Al shelves pressure 5– 10 mbar)
onto temp 100 ° C 110 min
onto temp 200 ° C 130 min
Time of rise onto 98 % voltage 230 V – ( shelves pressure 5– 10 mbar)
onto temp 100 ° C 170 min
onto temp 200 ° C 220 min
Heat radiation
at 100 ° C 370 W
at 200 ° C 750 W
Vacuum connection
feedthrough needle valve for inert gas or air 8 mm
measuring acces 40 mm
vacuum connection measuring DN 16 mm
Electric parameters – mains 50/ 60 Hz
nominal voltage 230 V
mains 50/ 60 Hz 1, 8 kW
current 7, 8 A
input in stand by mode 5 W
Max. permissible load of the shelves
inside the oven 65 kg/ oven
per tray 25 kg/ tray
* not measured
* * The heat in vacuum is transferred to the goods on the shelves by conduction in the shelves, therefore the mentioned temperature deviations are valid for temperatures on the surface of the shelves, there must be a perfect heat-conducting contact between the temperature sensors and the shelf surface. Goods placed on the shelves must also be in a perfect contact with the shelves, the goods temperature depends especially on their physical properties and on the contact with the shelf.