Detail MMM* DUROCELL, Ovens With Natural Air Convection 222 Litre, MC000915
Durocell 222 - Standard
Highly durable laboratory oven
202-liter chamber volume
Working temperature of from + 5° C over ambient temperature up to 125° C
Natural fine air flow in the chamber, noiseless operation
High-quality and safe drying, heating and tempering of laboratory materials in laboratories and industry; perfect anticorrosive protection of all chamber parts ( protective Epolon plastic) enabling wirk with acids and dases, acid and alkaline hydrolysis etc.
High operation comfort, precise temperature regulation and short times of temperature recovery in the chamber after the door opening
Standard model with a Fuzzy-Logic microprocessor control system; control elements situated on a foil keyboard, process information indicated on a LED display
Wide range of optional equipment and accessories
Output for a printer or PC connection
Validation possibility ( IQ , OQ)
Technical data
Interior of stainless steel material No. 1.4301
volume 222 l
depth 540 mm
width 540 mm
height 760 mm
Trays
standard equipment 2 pcs
storage area ( w × d) 520× 485 mm
racks 10 max. no
distance between guide rails 70 mm
Maximal weight of the load* )
inside the oven 70 kg/ oven
per tray 30 kg/ tray
Door
1 pcs
External dimensions ( including door and handle)
depth 790 mm
diameter of the air branch outer/ inner 52/ 49 mm
width 760 mm
height ( inclusive foots - N) 1090 ( N) mm
Package dimensions ( three layers carton)
depth 860 mm
width 830 mm
depth 1280 mm
Weight
gross 116 kg
nett 100 kg
Working temperature ( beginning of the regulation)
from 5 ° C above ambient temp. to: 125 ° C
Temperature accuracy according to DIN 12 880 T2, at working temperature with closed air flap and door
time variation cca ( ± ) 0, 3 ° C
space deviation cca ( ± ) % of the reached temperature 2 %
Time required to reach 100 ° C with closed air flap and voltage 230 V
50 min
Heat emission at 100 ° C
630 W
Air exchange speed at 100 ° C
5 per hour
Electricity – mains 50/ 60 Hz
max. power input 1, 8 kW
nominal voltage 230 V
current 7, 8 A
input in stand by mode 5 W
* ) Approx. 50 % of the tray area can be filled in a way a uniform air circulation is enabled inside the chamber.
Note: All technical data are related to 22 ° C ambient temperature and ± 10 % voltage swing ( if not specified) .
Tampilkan Lebih Banyak