Detail MMM* CO2CELL, Incubators With Natural Air Circulation And CO2 Atmosphere 190 Litre
CO2CELL 190 - Standard
Laboratory incubator – creates always the right atmosphere
Chamber volume 190 liters
Working temperature: From 5 ° C above ambient temperature up to 60 ° C
Unique system of the chamber and outer door heating, eliminating the need of a fan; ensures both uniform air distribution in the chamber and maximum relative humidity with absolutely dry inner walls
Suitable for high-quality and safe processing of microbiological samples both in health care ( IVF) and scientific laboratories
Control elements situated on a foil keyboard, process information indicated on a well-arranged LED display
Wide range of optional equipment and accessories
Output for a printer or PC connection
Validation possibility ( IQ , OQ)
Technical data
Inner dimensions
volume 190 l
depth 440 mm
height 632 mm
height 686 mm
Shelves ( trays) *
standard equipment 4 pcs
Number 8 max. pcs
External dimensios
depth 734 mm
height 765 mm
height 862 mm
Weight
gross 109 kg
nett 94, 5 kg
Temperature data
working temperature 5 ° C above ambient temperature to 60 ° C
Temperature deviations at 37 ° C max.
in time ± 0, 1 ° C
in time ± 0, 1 ° C
in space ± 0, 2 ° C
CO2 Control
control range 0, 2 - 20 % CO2
Control accurancy
in space ± 0, 2 %
measurement ± 0, 1 %
measurement ± 0, 1 %
Recovery rate
< = 0, 8 % / min
Required CO2pressure
0, 3-0, 7 / 5-10 bar / psi
Relative humidity
humidity dish volume 1, 5 l
Humidity level at 37 ° C, 5 ° C CO2
standard 95 % RH
higher 97 % RH
Electric parameters
max. output 380 ( high temperature dekontamination: 1800) W
Voltage ( 50/ 60 Hz) 230 ( + / - 10 % ) V
consumption at 37 ° C 0, 058 kWh
* ) Approx. 50 % of the tray area can be filled in a way a uniform air circulation is enabled inside the chamber.