Detail MMM* INCUCELL, Incubator With Natural Air Convenction, 55 Litre No.Cat: MC000722
INCUCELL 55 - Standard
Natural, precise – laboratory incubator
55-liter chamber volume
Working temperature of from + 5° C above ambient up to 99, 9° C
Noiseless operation and fine air flow in the chamber
High operation comfort, precise temperature regulation and short times of temperature recovery in the chamber after the door opening
High-quality and safe drying, heating and tempering of common laboratory materials in laboratories and industry; suitable for safe processing of microbiological cultures
Standard model with a Fuzzy-Logic microprocessor control system; control elements situated on a foil keyboard, process information indicated on a LED display
Wide range of optional equipment and accessories
Output for a printer or PC connection
Validation possibility ( IQ , OQ)
Technical data
Interior of stainless steel material No. 1.4301
volume 55 l
depth 370 mm
width 400 mm
height 350 mm
Tray
standard equipment 2 pcs
storage area ( w × d) 380× 335 mm
racks 4 max. number
distance between guide rails 70 mm
Maximal weight of the load* )
inside the oven 50 kg/ oven
per tray 20 kg/ tray
Door
1 pcs
External dimensions ( including door and handle)
depth 640 mm
diameter of the air branch outer / inner 52/ 49 mm
width 620 mm
height, incl. Feet ( N) and Rolls ( K) 680 ( N)
Package dimensions ( three layers carton)
depth 730 mm
width 700 mm
depth 880 mm
Weight
gross 66 kg
nett 55 kg
Working temperature ( beginning of the regulation)
from 5 ° C above ambient temp. to: 99, 9 ° C
Temperature accuracy according to DIN 12 880 T2 at working temperature with closed air flap and door
time variation cca ( ± ) < = 0, 2 ° C
space deviation cca ( ± ) < 0, 5 ° C
Time required to reach 37 ° C with closed air flap and voltage 230 V
49 min
Heat emission at 37 ° C
30 W
Air exchange speed at 37 ° C
5 per hour
Electricity - mains 50/ 60 Hz
max. power 0, 3 kW
nominal voltage 230 V
current 1, 3 A
power by stand by 5 W
* ) Approx. 50 % of the tray area can be filled in a way a uniform air circulation is enabled inside the chamber.
Note: All technical data are related to 22 ° C ambient temperature and + -10 % voltage swing ( if not specified) .